The advantages of performing cataract surgery through incisions as small as 1.5 mm include quicker healing, reduced amounts of induced astigmatism, a lesser likelihood of morphological changes to the cornea, and less postoperative inflammation.1-3 To implant currently available IOLs, however, surgeons often must widen the corneal incision to between 3.0 mm and 3.2 mm. A soft foldable IOL that may be implanted without requiring an enlarged incision is therefore an attractive development.
Three such lenses are currently available in Europe (Acrismart [Acritec GmbH, Berlin], Ultrachoice 1.0 [Thinoptx, Abingdon, Va], and Acriflex [Acrimed GmbH, Berlin]), but I have not found any of them entirely satisfactory in the postoperative course. In my experience with these IOLs, there is a higher rate of early posterior capsular opacification (PCO) during the first postoperative year compared with other lenses, as well as a higher incidence of anterior capsule opacifications and capsular shrinkage and phimosis.
Bausch & Lomb (Rochester, NY) recently developed the Akreos Micro-Incision Lens that may be injected using microcartridges through an incision of <2 mm. My group performed a pilot study to evaluate the performance of this IOL. Our objectives were to confirm that the lens can be implanted through an incision <2 mm, to evaluate two injection techniques, to assess the optical quality and intracapsular behavior of the lens, and to obtain preliminary information about the development of PCO with this IOL.4
ABOUT THE LENS
The Akreos Micro-Incision Lens is a single-piece, biconvex, hydrophilic acrylic IOL consisting of a central optic and four flexible haptics (Figure 1). The optic and haptics have square edges, and the haptics are angulated (approximately 11º), which helps to push the lens against the posterior capsule after its implantation in the capsular bag and thus increase its stability. The lens is available in three diameters 10.5 mm, 10.7 mm and 11 mm depending on the power, ranging from 10.00 D to 30.00 D. Like other IOLs in the Akreos line, the Akreos Micro-Incision Lens has a continuous 360º barrier edge to limit PCO. Table 1 provides additional characteristics of the lens, which is designed for implantation in the capsular bag of the posterior chamber after phacoemulsification. The IOL is designed for insertion with a 1.8 Viscoglide cartridge and the Viscoject Lens Injection System (both by Medicel AG, Widnau, Switzerland).
METHODOLOGY
Study design and subjects. The prospective, randomized, single-center study enrolled 19 patients who were 50 years of age or older and were scheduled to undergo cataract surgery. Subjects were randomized to in-the-bag implantation of an Akreos Micro-Incision Lens via either a standard inside-the-eye (in) or outside-the-eye (out) technique. These are described later.
Exclusion criteria included monocular status, traumatic cataract, intraocular surgery in the previous year, previous damage to the corneal endothelium, any ocular pathology other than a cataract, and participation in another clinical trial.
Surgery. After administering topical anesthesia, the surgeon created two microincisions of 1.5 mm in clear cornea, one in the superior-temporal region and the other at an angle of 90º to the first incision. Subsequently, he performed a continuous circular capsulorhexis, followed by extracapsular cataract extraction using bimanual phacoemulsification. The surgeon then injected a viscoelastic into the capsular bag through the first incision and implanted the IOL via either the in or the out technique.
The in technique involved inserting the Viscoglide cartridge through the first corneal incision into the anterior chamber and delivering the lens directly into the capsular bag. This technique requires an enlargement of the corneal incision after phacoemulsification from 1.5 mm to 2.3 mm. In the out technique, the surgeon positioned the Viscoglide cartridge outside the eye at the lip of the incision. After injection, the inferior haptic unfolds in the capsular bag, but the surgeon must position the superior haptics inside the capsular bag by means of a micromanipulator. This technique requires a calibration of the 2-mm incision.
RESULTS
Ten patients underwent the in technique versus nine for the out. The size of the corneal incision after the IOL's implantation was not significantly different from that before its insertion for either injection technique (Table 2). Figure 2 shows the postoperative visual acuities. The estimated A-constant required an adjustment to prevent unwanted refractive errors (Table 3), and Bausch & Lomb plans to recalculate the A-constant in order to improve the accuracy of IOL power calculations.
The Akreos Micro-Incision Lens was stable in the capsular bag with no significant decentration at the 15-day, 1-month, and 3-month postoperative examinations. Although insignificant, there was a slight shift toward the back of the eye in the first 2 weeks followed by a slight shift in the opposite direction during the next 2 months. The movement only occurred in seven patients and had no overall detrimental effect on visual acuity. PCO results will not be available until after the 12-month follow-up examination, but we have observed none in the nine patients who have reached 3 months' follow-up.
CONCLUSION
The pilot study indicates that the Akreos Micro-Incision Lens can be inserted through corneal incisions of <2 mm in length. The IOL has provided good visual acuity and remained centered over the pupil. The out technique was easier and faster to perform, and it required a smaller incision than the in technique.
My colleagues and I are planning to conduct a long-term monitoring program to confirm the performance of the Akreos Micro-Incision Lens, particularly with respect to its intracapsular stability and its limitation of PCO.
Thierry Amzallag, MD, is a surgeon at the Ophthalmic Institute of Somain in France. He states that he is a paid consultant for Bausch & Lomb. Dr. Amzallag may be reached at thierry.amzallag@institut-ophtalmique.fr or +33 327 998 853 .
Up Front | May 2006
The Akreos Micro-Incision Lens
A clinical evaluation of an IOL for microincisional cataract surgery.
Thierry Amzallag, MD