One of my colleagues visits local nursing homes to examine patients who are not easily referred to our hospital, mostly because of poor physical or mental health. From one such screening session, an elderly woman using a wheelchair was referred to me with poor vision in both eyes. She was a high myope and presented with a grade 4+ brown nuclear cataract in one eye. She was no longer able to watch television or read a newspaper or book. I promised to do the best I could to restore her vision. Unfortunately, an unexpected disaster arose—iatrogenic zonular dehiscence. However, I was able to correct the complication, and the patient achieved good quality of vision in the end.
In this case description with accompanying video clip (available at http://eyetube.net/v.asp?kozade), I explain my strategy for this iatrogenic zonular dehiscence disaster. Capsular folds during capsulorrhexis creation (Figure 1) are a clear indication of zonular weakness.
In any eye with suspected weak zonules, it is in the surgeon's best interest to create the largest feasible rhexis to reduce the stress on the zonules as much as possible. Additionally, the nucleus should be chopped into small pieces that are easily pulled into the anterior chamber, which will facilitate emulsification without inducing zonular stress. I know the theory of these maneuvers very well; however, in this case, I did not execute it accordingly. My failure to chop the nucleus into small pieces led to the iatrogenic disaster of 180° zonulolysis (Figure 2).
SURGICAL COURSE
What should be done in such a situation? In all cases
with a potential (partial) dropped nucleus, leave the
phaco tip in the eye to minimize fluid movements, which
are the major cause of nuclear material falling backward.
With the phaco tip in a still position, remove the second
instrument and inject any ophthalmic viscosurgical device
(OVD) to stabilize the position of the nucleus. Next, inject
a dispersive OVD between the nucleus and the vitreous.
My strategy is to create an absolute viscoelastic barrier to
prevent any nucleus drop as well as any direct contact with the vitreous. In the video and in Figure 3, you see profuse
amounts of Viscoat (Alcon Laboratories, Inc., Fort
Worth, Texas) injected underneath the nucleus.
VERY LOW FLUIDICS STRATEGY
In challenging and complicated cases such as zonular
dehiscence or posterior capsular rupture, I prefer to use
very low fluidics with torsional ultrasound, which I
described in CRST Europe's Second Annual Cataract
Complications issue (February 2007) in detail. Fluidics
settings (Table 1) should minimize pressure changes in
the anterior chamber, which are to be avoided because
high intraocular pressure has the tendency to induce
nucleus and/or vitreous movement. A moderate vacuum
setting, to avoid significant surge on occlusion
break, is recommended.
A very low aspiration flow rate (Figure 4) will reduce the speed of phacoemulsification but, more important, it will not aspirate too much of the visco-shield between the vitreous and the nucleus. With an adequate visco-shield in place, you can emulsify the nucleus safely in the anterior chamber. Prior to emulsification, the nuclear pieces should be brought up to the safe zone in the anterior chamber. This can be achieved with the help of the dispersive OVD, as illustrated in the video.
Torsional ultrasound enables the surgeon to emulsify nuclear material efficiently, regardless of density, at very low fluidics settings. There is virtually no chatter of nuclear chips, which could lead to unnecessary manipulation and increased risk for nuclear material to drop into the posterior segment. When using longitudinal ultrasound in cases with zonular dehiscence or posterior capsular rupture, reducing the ultrasound duty cycle significantly will reduce repulsion.
After all nuclear material has been emulsified, the protective visco-shield can be removed with bimanual vitrectomy. Diluted triamcinolone is employed to identify and remove any remaining vitreous.
CONCLUSION
This seemingly hopeless complicated case ended with a
positive outcome. The elderly woman decided to remain
aphakic as a +4.00 D hyperope, with excellent television and
reading vision. She was happy with her final result (Figure 5).
A very low fluidics strategy in the presence of a protective dispersive visco-shield provides the surgeon with a safe and controlled technique to successfully manage an iatrogenic posterior capsular tear or zonular dehiscence.
Khiun F. Tjia, MD, is an Anterior Segment Specialist at the Isala Clinics, Zwolle, Netherlands. Dr. Tjia states that he is a consultant to Alcon Laboratories, Inc., and Hoya. He is the Co-Chief Medical Editor of CRST Europe. Dr. Tjia may be reached at e-mail: kftjia@planet.nl.